Specialized - Radiation Counter ·GeigerMullerCounter

1B124

Metal Geiger-Müller counter intended for gamma-ray measurement in equipment and well-logging service. Rugged construction and field-oriented application make it suitable for subsurface radiation surveys and related instrumentation.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
A182
Base pins
-
Nominal heater
-

Description

1B124 — Metal Counter Thyrode Tube

The 1B124 is a sturdily constructed metal Geiger-Mueller counter tube, designated as a Thyrode. It is designed primarily for gamma ray measurement in portable survey instruments and well logging equipment. Positioned as an intermediate-sized device, the 1B124 fits between the smaller 1B85 Thyrode and the larger 1B125 Thyrode in its physical size class.

Mechanical Construction and Physical Dimensions

The 1B124 features a robust metal envelope designed to withstand demanding field environments. The wall of the tube is constructed from brass with a density thickness of 200 mg/cm². The tube has an active length of 11 inches, which defines its sensitive detection region.

The physical dimensions of the device are as follows:

  • Overall Length: 14 9/16 inches
  • Main Body Diameter: 0.750 inches
  • Base Diameters: 1 inch, 51/64 inch, and 15/16 inch

Base, Connections, and System Compatibility

The tube utilizes a standard RMA type A1-82 counter tube base, which is a coaxial base design. As a two-terminal device, its electrical connections are simplified into an anode and a cathode:

  • Anode (A): Connected to the center pin of the coaxial base.
  • Cathode (K): Connected to the outer shell/sleeve of the base.

This coaxial base configuration, combined with its nominal 900 V operating characteristics, allows the 1B124 to be interchangeable with several other counter tubes, including the 1B85 beta-gamma tube, the 1B106 mica window tube, and the 1B125 cosmic ray tube.

For physical integration into radiation detection equipment, the 1B124 is compatible with specific sockets and probes:

  • Compatible Sockets: Victoreen socket types 380 and 500.
  • Compatible Probes: Victoreen probe types 389 and 263-232.

Electrical Characteristics and Operating Ratings

The 1B124 operates as a high-voltage radiation detector with defined plateau and recovery characteristics. Its electrical performance parameters include:

  • Operating Voltage Range: 850 V to 950 V (with a nominal center point of 900 V).
  • Threshold Voltage: Maximum of 800 V.
  • Plateau Length: Minimum of 200 V.
  • Plateau Slope: Maximum of 3% per 100 V, measured across the voltage range of V0 = 800 V to 1000 V.
  • Recovery Time: 100 microseconds.
  • Background Count Rate: 200 counts per minute (c/m) when operated at V0 = 900 V.
  • Service Life: Rated for 10^9 counts when tested at a rate of 6000 c/m at V0 = 900 V. The end point of its useful life is defined when the plateau slope degrades to 10% per 100 V (measured from V0 = 850 V to 950 V).

Environmental Ratings

The 1B124 is rated to operate under a wide range of environmental conditions, making it suitable for both portable outdoor survey meters and deep well logging applications:

  • Ambient Temperature Range: Minimum of -10 °C to a maximum of 100 °C.
  • Relative Humidity: Maximum of 100%.

Socket description

RETMA/RMA A1-82 two-contact coaxial radiation-counter tube base family

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
A182
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 0

GeigerMullerCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1950 Victoreen

Published operating points and test conditions from this source.

SECTION 0

GeigerMullerCounter

Other

  • ThresholdVoltage

    800 V

    Max

  • Operating Voltage

    850 to 950 V

Characteristic

  • PlateauLength

    200 V

    Minimum

  • PlateauSlope

    30 %/100V

    Maximum

  • RecoveryTime

    100 us

  • Background

    200 counts/min

  • ActiveLength

    12 in