Skip to main content

Specialized - Radiation Counter ·GeigerMullerCounter

1B106

Mica-window Geiger-Müller beta counter designed for reduced operating voltage, high sensitivity, and long counting life. Its windowed construction suits laboratory and field measurement of beta-emitting sources.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
A182
Base pins
-
Nominal heater
-

Description

1B106 — Halogen-Quenched Beta Counter Tube

The 1B106 (also designated as a Thyrode) is a halogen-quenched, mica-window beta counter tube designed for laboratory and field beta radiation detection and measurement. Engineered for high reliability under demanding conditions, the tube features robust construction and is specifically designed for easy replacement in water-tight installations.

Construction and Physical Design

The 1B106 is constructed with a thin mica window having a density of 2 mg/cm² to allow the entry of beta particles into the active volume. The active length of the tube is 1.7 inches.

A key design feature of the 1B106 is its halogen quenching gas. The use of a halogen quencher eliminates both photosensitivity and static charge effects on the mica window, ensuring stable operation. Mechanically, the tube is built to withstand rugged environments, offering a high resistance to vibration and shock that is much greater than that of the 1B67 tube.

Physical Dimensions and Base Connections

The tube utilizes a coaxial base configuration designed to fit a Thyrode Socket No. 380 or Probe No. 263-232. The physical dimensions and connection details are as follows:

  • Base Diameter: 1.352 inches
  • Overall Length: 4-9/64 inches to 4-23/32 inches
  • Pin Diameter: 1/16 inch
  • Anode Contact: Located at the top terminal, featuring a 1-20 thread that is 1/4 inch deep
  • Cathode Contact: Formed by the bottom coaxial shell and pin

Electrical and Operating Characteristics

The electrical parameters of the 1B106 define its performance as a Geiger-Mueller detector. The tube operates with a maximum threshold voltage of 825 V. It features a minimum plateau length of 150 V, with a maximum plateau slope of 10% per 100 V evaluated over an operating voltage range of 800 V to 1000 V.

The key electrical and performance specifications include:

ParameterValue
Maximum Threshold Voltage825 V
Minimum Plateau Length150 V
Maximum Plateau Slope10% / 100 V (from 800 V to 1000 V)
Recovery Time200 µsec
Electrode Capacitance2.5 pF (µµF)
Background Count Rate40 c/m (at Vo = 900 V)
Useful Life10^10 counts (at 6000 c/m and Vo = 900 V)

Environmental Ratings and Operating Limits

The 1B106 is rated to operate reliably under a wide range of environmental conditions, making it suitable for field instruments as well as laboratory setups.

  • Operating Voltage Range: 850 V to 950 V
  • Ambient Temperature Range: -55 °C to 70 °C
  • Maximum Relative Humidity: 100%

Performance Curves and Pulse Characteristics

The operational behavior of the 1B106 is characterized by specific relationships between the operating voltage (Vo), the counting rate, and the charge generated per pulse.

  • Counting Rate vs. Operating Voltage: The counting rate curve exhibits a distinct plateau that begins near 800 V and extends past 1000 V. Within this plateau region, the counting rate levels off at approximately 6000 counts per minute.
  • Charge Per Pulse vs. Operating Voltage: The charge delivered per pulse increases linearly with the operating voltage. At an operating voltage of 800 V, the charge per pulse is approximately 4 x 10^-10 Coulombs, rising linearly to approximately 13 x 10^-10 Coulombs at 1000 V.

Socket description

RETMA/RMA A1-82 two-contact coaxial radiation-counter tube base family

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
A182
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 0

GeigerMullerCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1950 Victoreen

Published operating points and test conditions from this source.

SECTION 0

GeigerMullerCounter

Other

  • ThresholdVoltage

    825 V

    Max

  • Operating Voltage

    850 to 950 V

Characteristic

  • PlateauLength

    150 V

    Minimum

  • PlateauSlope

    10 %/100V

    Maximum

  • RecoveryTime

    200 us

  • Background

    40 counts/min

  • Life

    10000000000 counts

  • WindowMass

    2 mg/cm2

Capacitance

  • ElectrodeCapacitance

    2.5 pF