Specialized - Radiation Counter ·GeigerMullerCounter

1B67

Self-quenching Geiger-Müller counter for beta and gamma radiation detection and measurement. The gas-filled tube provides pulse-counting response for general radiation survey and laboratory instrumentation.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

1B67 Beta-Gamma Counter Tube

Overview

The 1B67 (also designated as the 1B67/VG-10A) is a self-quenching Geiger-Müller counter tube designed for the detection of beta and gamma radiation. Manufactured by The Victoreen Instrument Company, this specialized gas-discharge device is engineered to operate in the Geiger region to detect ionizing radiation events. It features a thin window for admitting beta particles, a robust mechanical flange connection, and a high-voltage coaxial electrode configuration.

Mechanical Construction and Physical Dimensions

The physical design of the 1B67 is optimized for radiation detection and secure mounting. The tube features a cylindrical main body with a thin window at one end and a mounting flange at the other.

Window Specifications

To allow beta particles to enter the active volume of the counter, the tube is constructed with a thin window:

  • Window Diameter: 1-1/8 inches minimum (1-1/8 inches nominal)
  • Window Thickness: 2.0 to 2.6 mg/cm²
  • Window Recess: 3/32 inches deep, featuring a 26° bevel

Physical Dimensions and Connections

The mechanical structure is defined by the following dimensions and electrical connection points:

  • Overall Height: 4-1/4 inches
  • Main Body Diameter: 1-3/16 inches
  • Body Length: 2-1/8 inches (excluding the anode pin and mounting flange)
  • Flange Diameter: 2.223 inches maximum
  • Flange Thickness: 7/16 inches

The electrical connections are integrated into the mechanical mounting system:

  • Anode Connection: A top pin with a maximum diameter of 0.255 inches and a length of 3/4 inches.
  • Cathode Connection: A mounting flange equipped with two holes tapped for #8-32 screws, positioned on a 1-13/16 inch Bolt Circle (B.C.).

Electrical and Radiation Detection Characteristics

The 1B67 operates as a self-quenching radiation detector. Its electrical parameters define its performance within a Geiger-Müller counter circuit.

Operating Voltages and Plateau

The tube requires a high direct-current (d.c.) potential to establish the electric field necessary for gas multiplication:

  • Geiger Threshold Voltage: 1100 V d.c. maximum
  • Nominal Operating Voltage: 1200 V d.c.
  • Plateau Length: 200 V minimum
  • Relative Plateau Slope: 5.0% maximum per 100 V

The plateau length of at least 200 V ensures stable operation even if the high-voltage power supply experiences minor fluctuations around the nominal 1200 V operating point. The plateau slope of no more than 5.0% per 100 V indicates that the count rate remains highly consistent across this voltage range.

Performance and Detection Metrics

  • Beta Detection Efficiency: 99.8%
  • Nominal Background Count: 45 counts per minute at the nominal operating voltage
  • Nominal Recovery Time: 300 µs
  • Static Electrode Capacitance: 3.3 µµF

The recovery time of 300 µs represents the period after an ionizing event during which the tube is insensitive to subsequent radiation, which is a critical parameter for calculating dead-time corrections in high-count-rate environments.

Operating Parameters and Environmental Ratings

The 1B67 is rated for operation under specific environmental conditions to ensure stability and prevent damage to the delicate window and gas fill:

  • Ambient Temperature Range: 0 to 100 °C

The self-quenching gas mixture inside the tube allows it to terminate the ionization discharge internally without requiring external quenching circuitry, simplifying the associated electronic design.

Technical Summary

The 1B67 is a self-quenching beta-gamma Geiger-Müller counter tube designed by The Victoreen Instrument Company. Operating at a nominal voltage of 1200 V d.c., it features a threshold voltage of 1100 V d.c. maximum, a minimum plateau length of 200 V, and a maximum plateau slope of 5.0% per 100 V. Mechanically, the tube is characterized by a 1-1/8 inch thin window (thickness of 2.0 to 2.6 mg/cm²), a top anode pin, and a cathode flange with two #8-32 holes on a 1-13/16 inch bolt circle. It provides a beta detection efficiency of 99.8%, a nominal background of 45 counts/min, and a recovery time of 300 µs, operating safely within an ambient temperature range of 0 to 100 °C.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 0

GeigerMullerCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1950 Victoreen

Published operating points and test conditions from this source.

SECTION 0

GeigerMullerCounter

Other

  • GeigerThresholdVoltage

    1100 V

    Max

  • Operating Voltage

    1200 V

    Nominal

Characteristic

  • PlateauLength

    200 V

    Minimum

  • RelativePlateauSlope

    5 %/100V

    Maximum

  • Background

    205 counts/min

    Max

Capacitance

  • ElectrodeCapacitance

    2.5 pF