Skip to main content

Specialized - Radiation Counter ·GeigerMullerCounter

1B68

Self-quenching Geiger-Müller counter for beta and gamma radiation detection and measurement. Its gas-filled pulse-counting operation suits general radiation survey, monitoring, and laboratory instruments.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

1B68 — Self-Quenched Beta and Gamma Counter Tube

The 1B68 (also designated as the 1B68/VG-13) is a self-quenched Geiger-Mueller counter tube designed for the detection of beta and gamma radiation. Registered under RMA release number 705 on October 28, 1948, the device was manufactured by The Victoreen Instrument Company of Cleveland, Ohio. It is engineered to provide highly efficient radiation detection within a robust, specialized physical envelope.

Mechanical Construction and Physical Dimensions

The 1B68 features a cylindrical construction optimized for radiation sensitivity and physical integration into detection equipment. The tube is terminated with flexible wire leads rather than a standard pin base, facilitating direct soldering or custom mounting configurations.

The physical dimensions and construction details of the tube include:

  • Overall Length: 8-1/8 inches maximum
  • Diameter: 25/32 inches maximum
  • Lead Length: 3-1/2 inches for both the anode and cathode terminals
  • Anode Length: 2-5/8 inches
  • Sensitive/Active Length: 3 inches minimum
  • Wall Thickness: 30 mg/cm²

The thin wall thickness of 30 mg/cm² is critical to its function, allowing beta particles to penetrate the active volume of the tube while maintaining structural integrity.

Electrical and Detection Characteristics

The 1B68 operates in the Geiger-Mueller region, utilizing a self-quenching gas mixture to terminate the ionization discharge after each detection event. The electrical and performance parameters under standard operating conditions are specified as follows:

ParameterSpecification
Nominal Operating Voltage900 VDC
Maximum Geiger Threshold Voltage850 VDC
Minimum Plateau Length200 V
Maximum Relative Plateau Slope3.0% per 100 V
Nominal Background Count Rate35 counts/min (at operating voltage)
Nominal Recovery Time100 µs
Beta Detection Efficiency99.8%
Static Electrode Capacitance2.2 µµF

The low static electrode capacitance of 2.2 µµF helps minimize the pulse rise time and overall charge per pulse. The nominal recovery time of 100 µs defines the resolving time limitations of the tube during high-count-rate applications. The plateau characteristics ensure stable operation, with a slope of no more than 3.0% per 100 V over a minimum range of 200 V above the threshold.

Environmental Ratings

The 1B68 is rated for operation across a defined thermal range. The maximum ratings specify that the ambient temperature must be maintained within the following limits:

  • Minimum Ambient Temperature: 0 °C
  • Maximum Ambient Temperature: 100 °C

Operating the tube outside of this 0 to 100 °C range may affect the quenching gas characteristics, plateau stability, or overall detection reliability.

Technical Summary

The 1B68 is a specialized, self-quenched counter tube designed by The Victoreen Instrument Company for beta and gamma radiation detection. Characterized by its nominal 900 VDC operating voltage, high beta efficiency of 99.8%, and a low background count rate of 35 counts per minute, the tube provides reliable performance in radiation measurement applications. Its physical design features a 30 mg/cm² wall, a minimum 3-inch active length, and 3-1/2 inch wire leads for anode and cathode connections, making it a distinct and highly optimized component of mid-twentieth-century radiation detection technology.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 0

GeigerMullerCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1950 Victoreen

Published operating points and test conditions from this source.

SECTION 0

GeigerMullerCounter

Other

  • GeigerThresholdVoltage

    850 V

    Max

  • Operating Voltage

    900 V

    Nominal

Characteristic

  • PlateauLength

    200 V

    Minimum

  • RelativePlateauSlope

    3 %/100V

    Maximum

  • Background

    105 counts/min

    Max

Capacitance

  • ElectrodeCapacitance

    2.2 pF