Skip to main content

Specialized - Radiation Counter ·GeigerMullerCounter

18520

Halogen-quenched Philips Geiger-Müller counter designed specifically for gamma-radiation measurement, with a stable counting plateau and long-life fill for monitoring and radiation-measurement instrumentation.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

18520 — Halogen-Quenched Geiger-Müller Radiation Counter Tube

The 18520 is a halogen-quenched Geiger-Müller radiation counter tube designed specifically for the measurement of gamma radiation. It features a robust metal cathode construction and is filled with a mixture of neon, argon, and a halogen quenching agent. The tube is characterized by its long expected operating life, stable plateau characteristics, and wide operating temperature range.

Construction and Physical Dimensions

The 18520 is constructed with a cylindrical cathode made of a chromium-iron alloy consisting of 27% chromium and 73% iron. This cathode has a wall thickness of 0.7 mm, an inside diameter of 20.8 mm, and an effective length of 140 mm.

Physically, the tube has a maximum overall length of 170 mm and an outer diameter of 22.2 mm. Electrical connections are made via flying leads with a length of 100 mm. The leads are color-coded for polarity:

  • Black wire: Cathode (k)
  • Red wire: Anode (a)

The total weight of the device is 75 g. The interelectrode capacitance of the tube is 4.5 pF.

Operating Characteristics and Electrical Parameters

At an ambient temperature of 20 °C, the 18520 exhibits the following electrical and operating parameters:

  • Starting Voltage (VignV_{\text{ign}}): Minimum 330 V, Maximum 345 V.
  • Operating Voltage (VbV_b): Minimum 375 V.
  • Anode Resistor (RaR_a): Two series-connected 2.7 MΩ resistors (totaling 5.4 MΩ).
  • Plateau Length: Minimum 100 V.
  • Plateau Slope: Typically 0.05 %/V, with a maximum limit of 0.15 %/V.
  • Expected Life: Minimum $5 \times 10^{10}$ counts, enabled by the self-healing properties of the halogen quenching agent.

Background Count Rates

The background count rate of the tube depends heavily on shielding:

  • Unshielded: 90 counts per minute.
  • Shielded (with 2 inches of lead [Pb] and 1/4 inch of aluminum [Al]): 40 counts per minute.

Environmental Limits

The 18520 is rated for operation under extreme environmental conditions. Its absolute limiting value for ambient temperature (tambt_{\text{amb}}) ranges from -55 °C to +75 °C.

Circuit Configuration and Operating Example

In a typical application circuit, the 18520 is configured with a grounded cathode. The anode is connected to the positive high-voltage supply (+Vb+V_b) through two series-connected 2.7 MΩ resistors. The output signal is extracted directly from the junction between these two series resistors. This configuration provides the necessary current limiting and signal decoupling for pulse detection.

Curve Characteristics and Dead Time Behavior

The performance of the 18520 can be understood through its characteristic curves:

  • Plateau Curve (Counts/sec vs. VbV_b): The tube displays a stable plateau starting around 360 V and extending up to approximately 500 V. Above 550 V, the count rate begins to rise steeply, indicating the end of the plateau region and the onset of continuous discharge.
  • Dead Time Curve (Dead time vs. VbV_b): The dead time of the tube varies dynamically with the applied operating voltage. At 360 V, the dead time starts at approximately 170 microseconds. It rises to a maximum of about 210 microseconds at around 460 V, before decreasing to approximately 120 microseconds as the voltage is increased to 600 V.

Technical Summary

The 18520 is a highly reliable, halogen-quenched Geiger-Müller tube optimized for gamma radiation detection. Featuring a chromium-iron cathode, a starting voltage of 330–345 V, and a minimum operating voltage of 375 V, it offers a stable plateau of at least 100 V with a very low slope (typically 0.05 %/V). With its wide operating temperature range of -55 °C to +75 °C and an expected life exceeding $5 \times 10^{10}$ counts, the 18520 is well-suited for robust radiation monitoring applications.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 1

GeigerMullerCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Gas-filled radiation-counter section; counter-specific electrical characteristics are retained in TUBE_SECTION_PARAMETER.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1999 Philips 0-D

Published operating points and test conditions from this source.

SECTION 1

GeigerMullerCounter

Construction

  • Filling

    Neon, argon and halogen quenching agent

Capacitance

  • Interelectrode capacitance

    4.5 pF

Characteristics

  • Starting voltage minimum

    330 V

  • Starting voltage maximum

    345 V

  • Plateau length minimum

    100 V

  • Plateau slope typical

    0.05 %/V

  • Plateau slope maximum

    0.15 %/V

  • Unshielded background

    90 counts/min

  • Shielded background

    40 counts/min

    2inPb;1/4inAl

  • Expected count life minimum

    50000000000 counts

Other

  • Operating voltage minimum

    375 V