Low-background halogen-quenched Geiger-Müller beta counter intended for use with an anticoincidence guard counter, enabling sensitive beta measurements with reduced cosmic-ray and environmental background.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

18516 — Low Background Beta Counter Tube

The 18516 is a halogen-quenched, end-window beta counter tube designed specifically for low-level radiation measurements. It is typically operated in combination with a cosmic-ray guard counter, such as the type 18518, to minimize background interference.

Construction and Physical Characteristics

The 18516 features a specialized construction optimized for detecting beta particles while maintaining low intrinsic background counts. The tube is filled with a mixture of neon, argon, and a halogen quenching agent.

Window and Cathode

The entry window is constructed from ferrochromium, with the following physical specifications:

  • Effective Diameter: 27.8 mm
  • Effective Area: 6.1 square centimeters
  • Thickness: 10 mg/cm²

The cathode is composed of 28% chromium and 72% iron. Its physical dimensions are:

  • Wall Thickness: 1.2 mm
  • Inside Diameter: 27.8 mm
  • Effective Length: 18 mm

Mechanical Dimensions and Capacitance

The physical envelope is compact, designed for integration into shielded counting assemblies.

  • Body Diameter: 29.9 to 30.9 mm
  • Flange Diameter: 33 to 34 mm
  • Maximum Body Height: 22 mm
  • Maximum Overall Height: 34 mm
  • Weight: 27 g
  • Capacitance: 1.3 pF

Electrical and Operating Characteristics

The electrical behavior of the 18516 is highly dependent on the selected anode resistance. Typical operating parameters are specified at an ambient temperature of 25°C under two distinct anode resistor configurations: 10 MΩ and 5 MΩ.

ParameterAnode Resistor of 10 MΩAnode Resistor of 5 MΩ
Starting Voltagemax 375 Vmax 375 V
Recommended Operating Voltage600 V600 V
Plateau Range500 – 750 V500 – 750 V
Maximum Plateau Slope0.03 %/V0.04 %/V
Maximum Dead Time (at 600 V)70 microseconds40 microseconds

The recommended operating voltage of 600 V is selected arbitrarily within the plateau, making it particularly suitable for anti-coincidence circuits.

Background Radiation Performance

When shielded with 10 cm of iron on the outside and 5 cm of mercury on the inside, the background count rates are extremely low:

  • Standard Operation: Maximum of 9 counts per minute.
  • Anti-Coincidence Operation: Maximum of 1.3 counts per minute when operated in an anti-coincidence circuit with a type 18518 guard counter.

Absolute Limiting Values

To prevent damage and ensure stable operation, the tube must be operated within the following absolute limits:

  • Maximum Operating Voltage: 750 V
  • Minimum Anode Resistor: 5 MΩ
  • Ambient Temperature Range: -50°C to +75°C

Circuit Integration and Mounting Requirements

Proper circuit design and physical mounting are critical to achieving the low-noise performance for which the 18516 is designed.

Recommended Circuitry

The recommended circuit configuration utilizes a split anode resistor network consisting of Ra1 and Ra2, where the values are proportioned such that Ra2 is equal to 10% of Ra1 (Ra2 = 0.1 * Ra1). The output signal is coupled via a capacitor connected directly to the junction between Ra1 and Ra2.

Mounting and Maintenance

  • Anode Connections: The physical connection between the anode pin and the anode resistor must be kept as short as possible.
  • Stray Capacitance: The capacitance between the anode connection and earth must be minimized to prevent degradation of the pulse characteristics and dead time.
  • Environmental Maintenance: The tube must be kept dry and thoroughly cleaned. This is essential to prevent leakage currents across the insulating surfaces, which could otherwise mimic radiation pulses or cause instability.

Performance Curves and Environmental Behavior

The operational characteristics of the 18516 exhibit specific trends across varying temperatures and voltages:

  • Starting Voltage vs. Temperature: The starting voltage remains highly stable and flat at approximately 340 V across a wide temperature range from -20°C to +80°C. At lower temperatures, specifically down to -60°C, the starting voltage decreases to approximately 300 V.
  • Plateau Characteristics: The tube exhibits a highly stable, flat plateau extending from 500 V to 850 V. This flat response is maintained for both the Ra = 5 + 0.5 MΩ and 10 + 1 MΩ split-resistor configurations.
  • Dead Time vs. Voltage: The dead time does not remain constant; it decreases non-linearly as the operating voltage is increased from 400 V up to 1000 V.

Technical Summary

The 18516 is a specialized halogen-quenched end-window beta counter tube designed for high-precision, low-background radiation detection. Featuring a ferrochromium window and a chromium-iron cathode, it provides a stable plateau from 500 V to 750 V with an extremely low plateau slope (0.03 %/V to 0.04 %/V depending on the anode resistance). When operated in an anti-coincidence configuration with a guard counter and appropriate shielding, its background count rate can be reduced to a maximum of 1.3 counts per minute. Safe and reliable operation requires a minimum anode resistance of 5 MΩ, minimal stray capacitance to earth, and strict adherence to cleanliness to prevent leakage currents.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 1

GeigerMullerCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Gas-filled radiation-counter section; counter-specific electrical characteristics are retained in TUBE_SECTION_PARAMETER.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1999 Philips 0-D

Published operating points and test conditions from this source.

SECTION 1

GeigerMullerCounter

Construction

  • Filling

    Neon, argon and halogen quenching agent

Capacitance

  • Interelectrode capacitance

    1.3 pF

Other

  • Anode resistor

    10 Mohm

  • Recommended anticoincidence operating voltage

    600 V

Characteristics

  • Starting voltage maximum

    375 V

  • Plateau minimum

    500 V

  • Plateau maximum

    750 V

  • Plateau slope maximum

    0.03 %/V

  • Dead time maximum

    70 usec

    Vp=600V

  • Shielded background maximum

    9 counts/min

  • Anticoincidence background maximum

    1.3 counts/min

    WithGuardCounter