Specialized - Radiation Counter ·GeigerMullerCounter

18513

Self-quenching mica-window Geiger-Müller counter for alpha and beta particles and very-low-energy photons. Its thin window and high-rate design support sensitive counting of weakly penetrating radiation.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

18513 — Self-Quenching Radiation Counter Tube

A self-quenching radiation counter tube featuring a thin mica window, designed for the detection of alpha and beta particles as well as low-energy photons.

Overview

The 18513 is a specialized, rare-gas-filled radiation counter tube designed for detecting alpha particles, beta particles, and low-energy photons. It is equipped with a thin mica window at one end to allow the entry of low-energy ionizing radiation. The tube is intended for use in laboratory, medical, and industrial tracer techniques, thickness measuring instruments, and general survey or laboratory instruments.

Construction and Physical Characteristics

The 18513 is constructed with a rare gas filling and a robust cathode wall. Its physical design is optimized for both protection and sensitivity:

  • Mica Window: The tube features a circular mica window with a diameter of 6.35 mm and an effective area of 31.7 mm². The density of the mica window ranges from 1.6 to 2.1 mg/cm², which is uniform for each individual counter.
  • Cathode Wall: The cathode wall has a thickness of 1 mm.
  • Enclosure and Accessories: The tube is delivered inside a protective aluminium container that is electrically connected to the cathode. The tube can remain inside this container during testing or when measuring gamma radiation and high-energy particles. A V-groove is provided on the container, allowing it to be cut with a pen knife to remove the counter tube when necessary. A removable rubber cap is also provided to protect the delicate mica window when the tube is not being used to measure low-energy radiation.

Physical Dimensions and Weight

The physical specifications of the 18513 vary depending on whether the tube is operated alone or within its protective aluminium container:

ParameterTube AloneIn Container
Maximum Length85.7 mm110 mm
Maximum Diameter14.2 mm37 mm
Weight18 g70 g

Electrical Capacitance

The electrical capacitance of the device is also affected by the presence of the outer aluminium container:

  • Tube alone: 1 pF
  • In container: 5 pF

Operating Characteristics and Environmental Ratings

The 18513 operates as a Geiger-Müller style counter with the following electrical and operational parameters:

  • Threshold Voltage: 575 V
  • Plateau Length: 250 V (with a minimum limit of 150 V)
  • Plateau Slope: Typically 0.06 %/V, with a maximum limit of 0.15 %/V (measured at 100 counts/sec)
  • Dead Time: Ranges from 70 to 40 µs, depending on the applied working voltage
  • Counting Rate: Maximum of 20,000 counts/sec
  • Background Noise: Maximum of 6 counts/min when shielded with 1/4" of aluminium (Al) and 2" of lead (Pb)
  • Ambient Temperature Range: -55 °C to +75 °C
  • Count Life Expectancy: Minimum of 10¹⁰ counts

Detection Efficiency and Performance

The detection capabilities of the 18513 for different types of radiation are documented across different specifications, showing variations in sensitivity and window stopping power:

Radium Source Sensitivity (Gamma/Radium)

  • Specification A: The tube exhibits a sensitivity of 1830 counts/min when exposed to a radium source of 2.39 ± 1% millicurie at 2.5 mR/hr (milli-röntgen per hour).
  • Specification B: The sensitivity is listed as 7210 counts/min under the same exposure conditions (2.39 ± 1% millicurie radium source at 2.5 mR/hr).

Particle Detection Thresholds

  • Beta Particles:
    • Under one specification, beta particles are counted if they arrive at the window with an energy greater than 30 keV.
    • Under an alternative specification, they are counted if they arrive at the window with an energy greater than 50 keV.
  • Alpha Particles:
    • Under one specification, the window stopping power is equivalent to 1.5 cm of air, and alpha particles are counted if they arrive at the window with an energy of approximately 2.5 MeV.
    • Under an alternative specification, the window stopping power is equivalent to 3 cm of air, and alpha particles are counted if they arrive at the window with an energy of approximately 5 MeV.

Circuit-Specific Operating Examples

To integrate the 18513 into radiation detection circuitry, typical operating configurations utilize the following component values:

Circuit Configuration 1

  • Cathode (k): Connected to ground via a 2.7 MΩ resistor.
  • Anode (a): Connected to the positive high-voltage supply via a 2.2 MΩ resistor.
  • Signal Coupling: Coupled from the anode via a 300 pF capacitor to a 330 Ω grid resistor (grid stopper) of an amplifier tube.
  • Grid Leak Resistor: 1 MΩ connected to ground.

Circuit Configuration 2

  • Cathode (k): Connected in series with a 2.7 MΩ resistor.
  • Anode (a): Connected to the positive high-voltage supply via a 2.7 MΩ resistor.
  • Signal Coupling: Coupled via a 300 pF capacitor to the grid of an amplifier tube.
  • Grid Circuit: Includes a 1 MΩ grid leak resistor to ground and a 330 Ω grid stopper resistor.

Technical Summary

The 18513 is a highly sensitive, self-quenching radiation counter tube featuring a 6.35 mm mica window and a rare gas filling. It is designed to detect alpha, beta, and gamma radiation across a wide temperature range of -55 °C to +75 °C, with an exceptional life expectancy of at least 10¹⁰ counts. With a threshold voltage of 575 V and a minimum plateau length of 150 V, the tube provides stable performance in laboratory, medical, and industrial applications. Its physical versatility is enhanced by a protective aluminium container and a detachable rubber cap, allowing it to be configured for either shielded high-energy measurements or open-window low-energy particle detection.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 1

GeigerMullerCounter

X-ray radiationInterelectrode capacitances
Max plate voltage
575 V

Published ratings and characteristics promoted semantically from the approved Philips source.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1957 Philips Receiving and Amplifying Tubes

Published operating points and test conditions from this source.

SECTION 1

GeigerMullerCounter

Characteristic

  • Starting / Threshold Voltage Minimum

    575 V

  • Plateau Length

    250 V

  • Minimum Plateau Length

    150 V

  • Plateau Slope Maximum

    0.15 %/V

    At published counting condition

  • Dead Time Maximum

    70 us

  • Background Count Maximum

    6 counts/min

    Published shielding condition

  • Mica Window Area

    31.7 mm2

  • Count Life Expectancy Minimum

    10000000000 counts

Other

  • Maximum Counting Rate

    20000 counts/s

  • Maximum Ambient Temperature

    75 C

Data set 02

1999 Philips 0-D

Published operating points and test conditions from this source.

SECTION 1

GeigerMullerCounter

Construction

  • Filling

    Rare gas

  • Mica window diameter

    6.35 mm

  • Mica window area

    31.7 mm2

  • Mica window areal density

    1.6-2.1 mg/cm2

  • Cathode wall thickness

    1 mm

Capacitance

  • Interelectrode capacitance

    1 pF

Characteristics

  • Threshold voltage

    575 V

  • Plateau length

    250 V

  • Plateau slope typical

    0.06 %/V

    At100CountsPerSecond

  • Plateau slope maximum

    0.15 %/V

    At100CountsPerSecond

  • Dead time minimum

    40 usec

    VoltageDependent

  • Dead time maximum

    70 usec

    VoltageDependent

  • Maximum counting rate

    20000 counts/sec

  • Shielded background maximum

    6 counts/min

    1/4inAl;2inPb

  • Expected count life minimum

    10000000000 counts