18511 — Proportional X-Ray Counter Tube

Overview

The 18511 is an organic quenched proportional X-ray counter tube designed for radiation detection and measurement. It features a mica side window and is filled with xenon gas combined with an organic quenching agent. The tube is optimized for detecting low-energy radiation, such as Manganese (Mn) K-alpha radiation, and operates in the proportional region below its Geiger threshold.

Construction and Physical Dimensions

The 18511 is constructed with a cylindrical body and a specialized side window for radiation entry. The internal gas filling consists of xenon at a pressure of 25 cm Hg, mixed with an organic quenching agent to stabilize the discharge process.

The physical dimensions and window specifications are detailed below:

  • Overall Length: 114 ± 3 mm
  • Maximum Diameter: 27 mm at the body and ends, increasing to 27.5 mm at the window collar
  • Cathode Inside Diameter: 21.5 mm
  • Effective Length: 67 mm
  • Mica Window:
    • Thickness: 2 to 2.5 mg/cm²
    • Dimensions: 7 mm × 18 mm
    • Area: 1.16 cm²

The tube has an anode-to-cathode capacitance of approximately 2 pF. To ensure reliable operation and prevent electrical leakage across the terminals, the tube must be kept dry and thoroughly cleaned.

Electrical and Operating Characteristics

The operating parameters of the 18511 are characterized at an ambient temperature of 25 °C using Manganese K-alpha radiation.

Voltage and Pulse Height

The tube operates within a high-voltage range, with the output pulse height depending directly on the applied anode voltage (V_b):

  • Operating Voltage (V_b): 1500 V to 1850 V. To maximize the operating life of the tube, the anode voltage should be maintained as low as possible within this range.
  • Geiger Threshold (V_b1): Greater than 1900 V. The tube must be operated below this threshold to remain in the proportional counting region.
  • Output Pulse Height (V_op):
    • At V_b = 1500 V, the output pulse height is 1 mV.
    • At V_b = 1850 V, the output pulse height rises to 100 mV.

Resolution and Background Noise

  • Energy Resolution (ΔP/P): ≤ 22%, where P represents the average pulse height and ΔP is the width of the pulse height distribution at half of the maximum value (Full Width at Half Maximum, or FWHM).
  • Integrated Background: When unshielded, the background count rate is approximately 15 counts per minute for pulses exceeding 50% of the average pulse height (P).

Absolute Maximum Ratings

Operating the tube outside its rated limits can cause permanent damage or severely degrade performance. The absolute limits for the 18511 are:

  • Maximum Anode Voltage (V_b): 1850 V
  • Ambient Temperature Range (t_amb): -50 °C to +75 °C

Circuit-Specific Operating Example

In a typical application circuit, the 18511 is integrated with an external amplifier stage to process the low-amplitude output pulses:

  • The anode of the counter tube is coupled directly to the grid of an amplifier tube.
  • A 0.1 MΩ resistor is placed in the grid/anode return path connected to the negative high-voltage supply (-V_b).
  • The associated amplifier tube utilizes a 2.2 kΩ resistor in its anode circuit to develop the amplified signal.

Technical Summary

The 18511 is a proportional X-ray counter tube featuring a 1.16 cm² mica side window, a xenon-organic gas filling at 25 cm Hg, and an anode-to-cathode capacitance of 2 pF. It operates in the 1500 V to 1850 V range, producing output pulses from 1 mV to 100 mV under recommended circuit conditions. With an energy resolution of ≤ 22% for Mn K-alpha radiation and a low unshielded background of 15 counts/min, it provides sensitive and resolved detection capabilities within an ambient temperature range of -50 °C to +75 °C.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 1

ProportionalCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Gas-filled radiation-counter section; counter-specific electrical characteristics are retained in TUBE_SECTION_PARAMETER.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1999 Philips 0-D

Published operating points and test conditions from this source.

SECTION 1

ProportionalCounter

Construction

  • Filling

    Xenon with organic quenching agent

  • Mica window area

    1.16 cm2

Capacitance

  • Interelectrode capacitance

    2 pF

Other

  • Operating voltage minimum

    1500 V

  • Operating voltage maximum

    1850 V

  • Absolute maximum operating voltage

    1850 V

Characteristics

  • Geiger threshold minimum

    1900 V

  • Voltage for 1 mV mean pulse

    1500 V

    MnKaRadiation

  • Voltage for 100 mV mean pulse

    1850 V

    MnKaRadiation

  • Energy resolution

    22 %

    MnKaRadiation

  • Integrated unshielded background

    15 counts/min

    PulseHeight>50%Mean