Skip to main content

Specialized - Radiation Counter ·GeigerMullerCounter

18510

Halogen-quenched Philips Geiger-Müller liquid-flow counter for beta-radiation measurement. Its construction permits radioactive liquid to pass through the sensitive counting region for continuous or sampled activity measurements.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

18510 — Halogen-Quenched Beta Radiation Liquid Flow Counter Tube

The 18510 is a specialized halogen-quenched Geiger-Müller counter tube designed for the detection and measurement of beta radiation in liquid flows. Utilizing a gas mixture of neon, argon, and a halogen quenching agent, the tube features an internal thin-wall glass flow-through design that allows liquids to pass directly through the active volume of the detector.

Mechanical Construction and Physical Properties

The 18510 is constructed with a thin-wall internal glass tube that serves as the conduit for the liquid flow being monitored. This internal glass tubing has a wall thickness of 30 mg/cm², an inside diameter of 5.5 mm, and an effective length of 36 mm. The cathode is fabricated from an alloy consisting of 28% chromium and 72% iron.

Physically, the tube has a maximum overall length of 92 mm and a maximum outer diameter of 24 mm. The total weight of the device is approximately 18 g. Because of the thin-wall internal glass tubing, the 18510 is mechanically fragile and must be handled with extreme care. Additionally, sudden changes in temperature must be avoided to prevent thermal shock and structural failure of the glass envelope.

Electrical and Operating Characteristics

At an ambient temperature of 25 °C, the 18510 exhibits the following electrical and operating parameters:

  • Anode-to-Cathode Capacitance: 4 pF
  • Ignition Voltage (VignV_{ign}): Maximum 375 V
  • Plateau Voltage Range (VplV_{pl}): 500 V to 650 V
  • Plateau Slope (SplS_{pl}): Maximum 0.07 %/V
  • Dead Time (τ\tau): Maximum 125 µs when operated at a supply voltage (VbV_b) of 550 V with a 5 MΩ anode resistor.
  • Background Count Rate (N0N_0): Maximum 15 counts per minute when the tube is shielded with 2 inches of lead (Pb) and 1/4 inch of aluminum (Al).

The operating voltage (VbV_b) can be selected arbitrarily within the plateau range to suit the requirements of the measurement system.

Recommended Operating Circuit and Absolute Limits

For proper operation, the 18510 is used in a grounded-cathode configuration. The anode is connected to the high-voltage supply (+Vb+V_b) through a series combination of two resistors: a primary anode resistor (RR) of 5 MΩ and a secondary tap resistor of 0.5 MΩ (representing a ratio of RR and 1/10 R1/10\,R). The pulse output is taken via a coupling capacitor connected to the junction between these two series resistors.

To ensure reliable operation and prevent damage, the following absolute limiting values must be respected:

ParameterLimit Value
Maximum Anode Voltage (VaV_a)650 V
Minimum Anode Resistor (RR)5 MΩ
Ambient Temperature Range (tambt_{amb})-50 °C to +75 °C
Maximum Internal Liquid Pressure (pp)120 cm Hg (absolute)

Temperature and Voltage Behaviors

The 18510 exhibits highly stable performance characteristics across its rated operating conditions:

  • Ignition Voltage vs. Temperature: The ignition voltage remains remarkably stable, holding at approximately 340 V to 345 V across a wide temperature range from -20 °C to +80 °C. At temperatures below -20 °C, the ignition voltage drops slightly, reaching approximately 300 V at -60 °C.
  • Dead Time vs. Operating Voltage: When utilizing the recommended 5 MΩ anode resistor with a 0.5 MΩ series tap, the dead time (τ\tau) remains relatively flat and stable. It stays within a narrow range of approximately 105 µs to 120 µs for operating voltages (VbV_b) spanning from 400 V to 700 V.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 1

GeigerMullerCounter

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Gas-filled radiation-counter section; counter-specific electrical characteristics are retained in TUBE_SECTION_PARAMETER.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1999 Philips 0-D

Published operating points and test conditions from this source.

SECTION 1

GeigerMullerCounter

Construction

  • Filling

    Neon, argon and halogen quenching agent

Capacitance

  • Interelectrode capacitance

    4 pF

Other

  • Anode resistor

    5 Mohm

  • Operating voltage minimum

    500 V

  • Operating voltage maximum

    650 V

  • Absolute maximum anode voltage

    650 V

  • Minimum anode resistance

    5 Mohm

  • Maximum liquid pressure

    120 cm Hg abs

Characteristics

  • Starting voltage maximum

    375 V

  • Plateau slope maximum

    0.07 %/V

  • Dead time maximum

    125 usec

    Vp=550V

  • Shielded background maximum

    15 counts/min

    2inPb;1/4inAl