Data set 01
1999 Philips 0-D
Published operating points and test conditions from this source.
GeigerMullerCounter
Construction
Filling
Neon, argon and halogen quenching agent
Cathode wall thickness
80-100 mg/cm2
Cathode inside diameter
4.8 mm
Cathode effective length
16 mm
Capacitance
Interelectrode capacitance
1 pF
Other
Anode resistor
10 Mohm
Operating voltage minimum
375 V
Operating voltage recommended maximum
450 V
WideAmbientRange
Characteristics
Starting voltage minimum
295 V
Starting voltage maximum
335 V
Plateau slope typical
0.07 %/V
Vp=450-550V
Plateau slope maximum
0.15 %/V
Vp=450-550V
Shielded background maximum
2 counts/min
2inPb;1/8inAl
Expected count life minimum
50000000000 counts