Skip to main content

Specialized - Radiation Counter ·GeigerMullerCounter

18506

Self-quenching Philips mica-window Geiger-Müller counter for alpha, beta, and gamma measurement. It combines a long operating plateau with halogen quenching and an enlarged sensitive window for radiation counting.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

18506 — Self-Quenching Geiger-Müller Counter Tube

A halogen-quenched, mica-window radiation counter tube designed for the detection and measurement of alpha, beta, and gamma radiation.

Overview

The 18506 is a self-quenching Geiger-Müller radiation counter tube designed for the detection and measurement of alpha, beta, and gamma radiation. It features a thin mica window at one end to permit the entry of low-energy alpha and beta particles, while its robust metal cathode and halogen-quenched gas filling ensure long-term stability and an exceptionally long operating life.

Construction and Physical Dimensions

The 18506 is constructed with a cylindrical metal cathode and a thin mica window.

  • Cathode: Made of an alloy consisting of 28% chromium and 72% iron. It has a wall thickness of 1.3 mm and an inside diameter of 27.8 mm. Depending on the production version, the effective cathode length is specified as 26 mm, 36 mm, or 37 mm.
  • Mica Window: Positioned at the front of the tube, the window has an effective diameter of 27.8 mm, corresponding to an effective area of 6 cm² (or 6.1 cm² in some versions). The window thickness is extremely low, ranging between 2.5 and 3.5 mg/cm², which minimizes attenuation of incoming radiation.
  • Gas Filling: The tube is filled with a mixture of Neon (Ne) and Argon (Ar), along with a halogen quenching agent that automatically terminates the ionization discharge after each pulse.
  • Physical Dimensions:
    • Body Diameter: 30.4 mm (increasing to a maximum of 33.5 mm at the window flange).
    • Body Length: 36 mm (excluding the anode pin and a 2.2 mm window recess).
    • Anode Pin: Measures 12 mm in length and 1 mm in diameter, spaced 5 mm from the main body.
  • Interelectrode Capacitance: The capacitance of the tube is rated at 2 pF (or up to 3.5 pF in later versions).

Electrical and Operating Characteristics

The 18506 operates in the Geiger-Müller region, where the output pulse amplitude is independent of the primary ionization. Typical operating parameters at an ambient temperature of 25 °C include:

  • Starting Voltage (VaV_a / VignV_{ign}): Minimum of 325 V and maximum of 375 V. The starting voltage exhibits a temperature coefficient of approximately 0.5 V/°C.
  • Operating Voltage (VaV_a / VbV_b): The tube can be operated arbitrarily within its plateau region, with a minimum recommended supply voltage (VbV_b) of 450 V in later specifications.
  • Plateau Characteristics: Measured at a count rate of 100 counts/second with a 10 MΩ anode resistor:
    • Beginning of Plateau: Maximum of 475 V.
    • Plateau Length: Typically 300 V (with a minimum limit of 250 V or 300 V depending on the specific operating conditions and version).
    • Plateau Slope: Typically 0.01 %/V, with a maximum limit of 0.02 %/V (specifically evaluated over the range of 450 V to 750 V).
  • Dead Time: The maximum dead time is 225 µs when measured at 100 counts/second with a 10 MΩ anode resistor. Typically, the dead time is around 180 µs at an operating voltage of 450 V.
  • Background Radiation: When shielded with 2 inches of lead (Pb) and 1/8 inch of aluminum (Al), the background count rate is extremely low, with a maximum of 25 to 40 counts per minute (impulses/min).
  • Expected Lifetime: The halogen quenching agent prevents degradation of the gas mixture, yielding an expected operating life of at least $5 \times 10^{10}$ counts.

Absolute Maximum Ratings

To prevent damage and ensure reliable operation, the following absolute limits must be observed:

  • Minimum Anode Resistor (RaR_a): 2 MΩ.
  • Ambient Temperature Range (tambt_{amb}): Minimum of -55 °C (or -50 °C in some versions) to a maximum of +75 °C.

Circuit and Mounting Recommendations

Proper electrical and mechanical installation is critical to prevent spurious pulses and leakage currents:

  • Mounting: A low-capacity mounting configuration with short connections is required. The connection between the anode and the anode resistor must be kept as short as possible, and the stray capacitance between the anode connection and ground (earth) must be minimized.
  • Maintenance: The tube envelope must be kept dry and thoroughly clean. Any surface contamination can lead to leakage currents, which degrade performance or cause false counts.
  • Measuring Circuit: The recommended series anode resistor (RR) is 10 MΩ. The output signal can be extracted via a coupling capacitor in two ways:
    1. Directly from the anode.
    2. From the junction of the 10 MΩ anode resistor and an auxiliary 1/10R1/10 R (1 MΩ) resistor connected to the positive high-voltage supply.

Qualitative Performance Curves

The operational behavior of the 18506 is characterized by several key relationships:

  • Count Rate vs. Anode Voltage: The discharge threshold begins near 350 V. A highly stable, flat plateau is maintained from approximately 400 V to 700–750 V, where the count rate remains steady (holding near 100 counts/second). Above 750 V, the count rate rises sharply as the tube enters the continuous discharge region, reaching approximately 120 counts/second at 900 V.
  • Dead Time vs. Anode Voltage: At the onset of conduction near 320–360 V, the dead time is approximately 110 µs. As the voltage increases, the dead time rises to a broad peak of approximately 180–190 µs between 450 V and 600 V, before gently declining to about 120–175 µs at 800–900 V.
  • Starting Voltage vs. Ambient Temperature: The starting voltage is highly stable at approximately 350 V across a wide temperature range from -20 °C to +80 °C. Below -20 °C, the starting voltage drops progressively, reaching approximately 290 V at -60 °C.

Technical Summary

The 18506 is a halogen-quenched, self-quenching Geiger-Müller counter tube featuring a 27.8 mm effective diameter mica window (2.5 to 3.5 mg/cm² thickness) and a chromium-iron cathode. Operating with a starting voltage between 325 V and 375 V, it provides a wide, stable plateau of at least 250 to 300 V with a very low slope (typically 0.01 %/V to 0.02 %/V). Designed for alpha, beta, and gamma radiation detection, the tube requires a 10 MΩ anode resistor, low-capacity mounting, and clean, dry operating conditions to achieve its rated lifetime of over $5 \times 10^{10}$ counts.

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
-
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 1

GeigerMullerCounter

X-ray radiationInterelectrode capacitances
Max plate voltage
475 V

Published ratings and characteristics promoted semantically from the approved Philips source.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1957 Philips Receiving and Amplifying Tubes

Published operating points and test conditions from this source.

SECTION 1

GeigerMullerCounter

Characteristic

  • Starting / Threshold Voltage Minimum

    325 V

  • Starting Voltage Maximum

    375 V

  • Plateau Length

    300 V

  • Beginning of Plateau Maximum

    475 V

  • Plateau Slope Maximum

    0.02 %/V

    At published counting condition

  • Dead Time Maximum

    225 us

  • Background Count Maximum

    40 counts/min

    Published shielding condition

  • Count Life Expectancy Minimum

    50000000000 counts

Other

  • Maximum Ambient Temperature

    75 C

Data set 02

1999 Philips 0-D

Published operating points and test conditions from this source.

SECTION 1

GeigerMullerCounter

Construction

  • Filling

    Neon, argon and halogen quenching agent

  • Mica window effective diameter

    27.8 mm

  • Mica window area

    6 cm2

Capacitance

  • Interelectrode capacitance

    2 pF

Characteristics

  • Starting voltage minimum

    325 V

  • Starting voltage maximum

    375 V

  • Beginning of plateau maximum

    475 V

  • Plateau length minimum

    250 V

  • Plateau slope typical

    0.01 %/V

  • Plateau slope maximum

    0.02 %/V

  • Shielded background maximum

    40 counts/min

    2inPb;1/8inAl

  • Expected count life minimum

    50000000000 counts