Data set 01
1957 Philips Receiving and Amplifying Tubes
Published operating points and test conditions from this source.
GeigerMullerCounter
Characteristics
Starting Voltage
900 V
Operating Voltage
1050 V
Capacitance
2 pF
Specialized - Radiation Counter ·GeigerMullerCounter
Philips self-quenching Geiger-Müller radiation counter intended for ionizing-radiation measurement. Manufacturer data specify counting plateau, background, timing, life, and construction characteristics for laboratory instrumentation.
The 18500 is a non-self-quenching, argon-hydrogen filled Geiger-Müller radiation counter tube. It is designed for the detection and measurement of ionizing radiation, including X-rays, gamma rays, and neutrons. Because it is a non-self-quenching design, it requires external circuitry to terminate the ionization discharge after each event.
The 18500 features a robust coaxial construction optimized for stable radiation detection. The cathode is fabricated from a specialized alloy consisting of 28% chromium and 72% iron.
Electrical connections to the tube are made via dedicated anode and cathode terminals:
To ensure proper operation and minimize parasitic effects, the tube requires a low-capacity mounting configuration utilizing short electrical connections. Additionally, the tube must be kept dry and thoroughly cleaned to prevent the formation of leakage currents across its surface.
The electrical behavior of the 18500 is characterized by a high operating voltage and a well-defined plateau region. The tube exhibits an internal electrical capacitance of 2 pF.
The tube begins to register counts at a starting voltage of 900 V. The plateau region, where the count rate remains relatively independent of the applied voltage, begins at 950 V and extends for a length of 150 V. Within this plateau, the slope is extremely flat, with a maximum change of 0.02% per Volt.
The recommended operating conditions are structured as follows:
When operated under specified quenching conditions, the tube has a dead time of 75 µs. It exhibits high sensitivity and efficiency, with a documented cosmic ray efficiency of 98.5%. The expected operational lifetime of the tube is 10¹⁰ counts.
When shielded with 2 inches of lead (Pb) and 1/8 inch of aluminum (Al), the background count rate is kept to a maximum of 20 counts per minute.
Because the 18500 is a non-self-quenching tube, it cannot terminate the gas discharge on its own. It must be operated in conjunction with an external quenching circuit. Specifically, a Neher-Harper quenching circuit must be utilized to quench the discharge and reset the tube for subsequent events.
The output of the tube and its quenching circuit must be connected to an external amplifier and scaler system. This processing system must meet the following minimum specifications:
The 18500 is designed for the measurement of X-ray, gamma, and neutron radiation.
For the specific application of thermal neutron detection, the tube must be wrapped in a cadmium foil with a thickness of approximately 0.5 mm. The cadmium foil acts as a converter, interacting with thermal neutrons to produce secondary radiation that can be readily detected by the argon-hydrogen filled active volume of the tube.
| Parameter | Specification |
|---|---|
| Tube Type | Geiger-Müller counter tube (non-self-quenching) |
| Gas Filling | Argon-hydrogen |
| Cathode Composition | 28% Cr, 72% Fe |
| Capacitance | 2 pF |
| Starting Voltage | 900 V |
| Operating Voltage Range | 1000 V to 1100 V (1050 V typical) |
| Plateau Length | 150 V (starting at 950 V) |
| Plateau Slope | Max. 0.02% per V |
| Dead Time | 75 µs |
| Cosmic Ray Efficiency | 98.5% |
| Expected Life | 10¹⁰ counts |
| Required Quenching Circuit | Neher-Harper |
General specifications
Properties common to every published source and operating point.
Electrical assignments shared by every tube using this base designation.
No verified pinout is recorded for this base designation.
1 physical section, each with unique specifications.
Section 1
No fixed ratings are recorded for this section.
Radiation-counter discharge tube.
Source-dependent data
Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.
Data set 01
Published operating points and test conditions from this source.
Starting Voltage
900 V
Operating Voltage
1050 V
Capacitance
2 pF
Data set 02
Published operating points and test conditions from this source.
Capacitance
2 pF
Counter tube
Cosmic ray efficiency
98.5 %
Published counter efficiency
Operating voltage minimum
1000 V
Geiger-Muller operation
Operating voltage maximum
1100 V
Geiger-Muller operation