Data set 01
1953 Amperex
Published operating points and test conditions from this source.
GeigerMullerCounter
Other
Starting voltage maximum
625 V
Characteristics
Maximum counting rate
830 counts/s
Dead time approximate
200 us
Specialized - Radiation Counter ·GeigerMullerCounter
Based version of the neon-filled self-quenching Geiger-Müller counter for beta detection, using a thin mica window and broad operating plateau for general radiation-counting instrumentation.
The 100NB is a permanent-sensitivity, thin-window radiation counter tube designed specifically for beta detection applications. Featuring a stainless steel cathode, a neon gas filling with a quenching admixture, and a thin mica window, the tube is engineered for stable radiation measurement under a nominal operating potential of 700 V DC.
The 100NB is constructed with a robust mechanical envelope and specialized window materials to facilitate the entry and detection of beta particles.
The tube incorporates a thin mica window at its entry port. The physical properties of this window are tightly controlled to minimize attenuation of beta radiation:
The cathode structure is fabricated entirely from stainless steel, ensuring long-term stability and resistance to degradation. The physical dimensions of the effective cathode are:
The tube is designed with a standard plug-in base for secure mounting and electrical connection:
The 100NB operates as a self-quenched Geiger-Müller style counter tube, utilizing a neon gas filling combined with a quenching admixture. This chemistry provides a stable plateau and prevents continuous discharge after an ionization event.
The 100NB is designed to operate reliably under demanding environmental conditions and offers an exceptionally long service life.
The 100NB shares identical electrical characteristics, gas filling, and window dimensions with the 100N counter tube. Both models operate at 700 V DC, feature a 200-volt plateau, and have a 200-microsecond dead time. However, they differ in physical packaging and base configuration: the 100N has maximum overall dimensions of 1-1/2 by 3-3/4 inches, whereas the 100NB is longer and narrower, measuring 1-3/8 by 4-11/32 inches, and is specifically equipped with a 4-pin base.
Additionally, the 100NB can be compared to the 100C and 100CB models in the same size class. While the 100NB uses a neon-based filling and operates at 700 V DC for beta detection, the 100C and 100CB utilize an argon-based filling with a higher operating voltage of 1200 V DC, a plateau in excess of 300 volts, and are rated for both Beta and X-Ray applications.
General specifications
Properties common to every published source and operating point.
Electrical assignments shared by every tube using this base designation.
No verified pinout is recorded for this base designation.
1 physical section, each with unique specifications.
Section 1
No fixed ratings are recorded for this section.
Counter-tube source does not publish ordinary active-electrode promoted fields; counter performance is retained in parameters.
Source-dependent data
Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.
Data set 01
Published operating points and test conditions from this source.
Starting voltage maximum
625 V
Maximum counting rate
830 counts/s
Dead time approximate
200 us
Data set 02
Published operating points and test conditions from this source.
Filling
Neon + quenching admixture
Operating voltage
700 V
DC
Window thickness
0.0005 in
MicaWindow
Window areal density
3.5 mg/cm2
MicaWindow
Effective mica-window diameter
1-3/32 in
Cathode dimensions
1-1/2 long x 1-3/16 OD x 3/32 wall
Maximum overall tube dimensions
1-3/8 x 4-11/32; 4-pin base
Radiation application
Beta
Operating temperature range minimum
-55 degC
Operating temperature range maximum
75 degC
Plateau length
in excess of 200 V
Plateau slope
5% to 10% per 100 V
Dead time
200 us
Approximate
Maximum shielded background
50 counts/min
Shielded by 2 inches of lead