Skip to main content

Specialized - Gauge Tube ·Gauge

7677

Thermocouple vacuum gauge using filament heating and gas-dependent thermal transfer, intended for roughly 1 to 10^-2 mm Hg pressure measurement.

Directly heatedCold cathodeCommon cathodeDissimilarSpecialized
Sections
1
Base
-
Socket code
-
Base pins
-
Nominal heater
-

Description

Westinghouse March 15,1961 THERMOCOUPLE VACUUM GAUGE TYPE 7677 The 7677 features a thermocouple heated by a filament, The internal gas pressure detennines the heat transfer characteristics and thereby the output of the thermocouple. The type 7677 is useful for pressure measureinents in the range from 1 to 10~2 millimeters of mercury. ELECTRICAL: OPERATIONAL: Thermocouple Materials . ... ......... Nichrome 8 Advance Heater Current. . . . .. oo v v o v v TR 25 Ma, Heater Resistance ¢t 25°C ... .. e e e 65 Chms Thermocouple Output: Thermocouple Resistance 0t 25°C ., ... ... ...... 10 O4me At 0= mm Hg Pressure, « v v v o v e e e e 0 mV¥. MECHANICAL: MOUNHIG POSTHION < o o v vt it vt v tn e s snnns Vertical Note: Other thermocouple gauge tubes, similar to type 7677 but Base. .« ..l it i i et et e JEDEC A4-9 having different sizes ond materiols of tubulation, ars Tubulation: available under different numbers, Diometar . . v i ittt i e s e e e e e, 3/8" Material . . . . ..ttt it e e, Nenex, C

General specifications

Tube and section characteristics

Properties common to every published source and operating point.

Heater and base

Base
-
Socket code
-
Base pins
-
Nominal heater voltage
-
Nominal heater current
25 mA
Parallel heater voltage
-
Parallel heater current
-
Series capable
No

Terminal pinout

Electrical assignments shared by every tube using this base designation.

No verified pinout is recorded for this base designation.

Section ratings

1 physical section, each with unique specifications.

Section 1

Gauge

X-ray radiationInterelectrode capacitances

No fixed ratings are recorded for this section.

Source-derived physical/electrical section.

Source-dependent data

Operating conditions by source

Each block preserves the test conditions and published values from one reference. Values from different sources are intentionally kept separate.

Data set 01

1947 Westinghouse

Published operating points and test conditions from this source.

SECTION 1

Gauge

Characteristic

  • Heater Resistance ¢t

    65

    Heater Resistance ¢t 25°C ... .. e e e 65 Chms Thermocouple Output:

  • Thermocouple Resistance

    0

    Thermocouple Resistance 0t 25°C ., ... ... ...... 10 O4me At 0= mm Hg Pressure, « v v v o v e e e e 0 mV¥.