1947 — Pirani Vacuum-Gauge Tube
The 1947 is a specialized Pirani-type vacuum-gauge tube designed for the determination of gas pressures in vacuum systems and vacuum enclosures. Operating on the principle of thermal conductivity, the tube utilizes a heated filament whose resistance varies with the pressure of the surrounding gas.
Physical Construction and Dimensions
The 1947 is constructed with a T-9 glass bulb made of Corning Code 001 Lead soft glass. It features a tubulation at the top of the bulb for direct connection to the vacuum system under test. The tubulation diameter is specified as either 7/32 inch or 3/8 inch.
The physical dimensions and mounting specifications of the tube include:
- Base: Small-Shell Small 4-Pin or Small 4-pin base.
- Mounting Position: Any.
- Maximum Overall Length: 7-3/16 inches to 7-9/16 inches (including the tubulation).
- Maximum Diameter: 1-3/16 inches.
- Pin-to-Bulb-Top Length: Approximately 5-3/16 inches, measured from the end of the base pins to the bulb-top line as determined by a ring gauge of 1/2 inch inside diameter.
Filament and Calibration Characteristics
The sensing element of the 1947 is a filament made of Platinum Iridium. Under operating conditions, the filament characteristics are:
- Nominal Voltage: Approximately 10 V DC.
- Filament Current: 70 to 100 mA (0.07 to 0.1 A), which varies depending on the gas pressure inside the envelope.
- Filament Resistance: 135.8 ohms under a high vacuum better than 3 x 10^-5 mm of mercury (Hg).
To ensure interchangeability and accuracy, each tube features a series filament-calibrating resistor (R) located directly within the base of the tube. Each individual tube is calibrated to establish a total resistance of exactly 135.8 ohms under a vacuum better than 3 x 10^-5 mm of Hg.
Pinout Configurations
The 1947 uses a small 4-pin base. Depending on the specific production run or specification standard, two terminal connection configurations are documented (viewed from the bottom of the base):
Configuration A
- Pin 1: Filament
- Pin 2: Filament
- Pin 3: No Connection (NC)
- Pin 4: Internal Connection (IC) — Do Not Use
Configuration B
- Pin 1: Internal Connection (IC)
- Pin 2: Resistance element (R)
- Pin 3: No Connection (NC)
- Pin 4: Resistance element (R)
In both configurations, Pin 3 remains unused, and care must be taken not to connect external circuitry to the designated Internal Connection pin.
Absolute Maximum Ratings
To prevent damage to the Platinum Iridium filament and ensure stable calibration, the operating conditions must not exceed the following absolute limits:
- Maximum Filament Voltage: 16 V
- Maximum Ambient Temperature: 60 °C
Bridge Circuitry and Calibration
The 1947 is operated within a Pirani gauge bridge circuit. A typical bridge configuration utilizes the following components:
- Power Supply: 12 V
- Resistors:
- R1: 50 Ohms (adjustable)
- R2: 25 Ohms (adjustable)
- R3 + Meter: 15 Ohms
- R4, R5: 10 Ohms each
- R6: 120.7 Ohms
- R7: 135.8 Ohms
- Switch (S): A selector switch with positions 1 and 2.
Calibration Procedure
- Set the selector switch S to position 2. Adjust the variable resistor R2 until the meter reads exactly 2.5 milliamperes.
- Set the selector switch S to position 1. With dry air at atmospheric pressure inside the 1947 tube, adjust the variable resistor R1 until the meter reads exactly 5.0 milliamperes.
- Without making any further adjustments, keep the switch S in position 1. The gauge is now calibrated and ready for vacuum measurement.
Pressure Sensitivity and Calibration Curve
The 1947 is capable of measuring gas pressures across a wide range, with its sensitivity characteristics defined as follows:
- Useful Sensitivity Range: 1.5 to less than 0.01 mm of Hg (equivalent to 1500 to less than 10 microns of Hg).
- Greatest Sensitivity Range: 0.5 to 0.01 mm of Hg (equivalent to 500 to 10 microns of Hg).
When calibrated with dry air using the standard bridge circuit, the relationship between the linear meter deflection (0 to 5 milliamperes) and the logarithmic gas pressure (0.01 to 3.0 mm of Hg) exhibits the following typical characteristics:
- At a high vacuum of 0.01 mm of Hg, the meter deflection is approximately 0.25 mA.
- At a pressure of 0.1 mm of Hg, the meter deflection rises to approximately 1.8 mA.
- At a pressure of 1.0 mm of Hg, the meter deflection reaches approximately 4.5 mA.
- As the pressure increases further, the curve rises extremely steeply as it approaches a meter deflection of approximately 4.7 mA. This represents the upper limit of pressure sensitivity for this specific bridge configuration. For measurements in microns, the values in mm of Hg are multiplied by 1000.